Temperature measurements provide 3-D ultrasound maps rapidly

SEP 24, 2018
A new thermographic method for visualizing ultrasound fields allows rapid scanning of volumes to produce 3-D maps of the sound intensity.
Temperature measurements provide 3-D ultrasound maps rapidly internal name

Temperature measurements provide 3-D ultrasound maps rapidly lead image

There are several techniques for visualizing ultrasound fields. One involves scanning a hydrophone, an underwater microphone, through a set of points. Another approach relies on acousto-optical effects, which are based on a change in the optical refractive index by pressure (i.e., sound) waves. Other techniques such as Doppler velocimetry have also been used, but all of these methods are slow or require specialized optical setups.

In an article by Melde et al., the authors describe a new method for producing 3-D maps of an ultrasound field using thermography. The measurement device consists of an open cylindrical cup that has a thin membrane across the bottom surface. A thermal camera mounted above the cup scans through various positions with an automated stage.

The membrane is made of a sound-absorbing material; the authors investigated two different types of membranes. Both membrane types convert absorbed sound waves to heat, which leads to a local rise in temperature. The camera records the detected temperature distribution across the membrane. Calibration of the membrane and camera allows conversion from temperature to sound pressure amplitude in the water tank below the cup.

Initial measurements show good agreement between experimental results and a simple model for membrane heating, and the data show the expected quadratic dependence of temperature on sound pressure. While the results appear to be not as sensitive or accurate as those gathered through hydrophone measurements, the data are collected much more quickly. A full 3-D image of a sound profile in a 54-cubic-centimeter volume was produced in just 33 seconds with this set-up – this is orders of magnitude faster than a hydrophone could achieve.

Source: “Fast spatial scanning of 3D ultrasound fields via thermography,” by K. Melde, T. Qiu, and P. Fischer, Applied Physics Letters (2018). The article can be accessed at https://doi.org/10.1063/1.5046834 .

Related Topics
More Science
/
Article
A simpler computational technique can provide qualitative assessments of target asymmetries.
/
Article
Simulations show that pulsations with intermediate frequency are best at mobilizing stuck droplets, a finding that could help with the transport of fluids in subsurface reservoirs and other porous materials.
/
Article
Open-source framework package constructs accurate and computationally efficient MLIPs without relying on an extensive prelabeled dataset
/
Article
Higher-order interactions are often found to be detrimental to synchronization, but a new study shows that, when weak, they can work together with pairwise interactions to enhance it.